K22680 Grease Mobility Test Apparatus

Test Method
Determines the resistance of lubricating grease to flow under prescribed conditions. Mobility is measured in grams per second by pumping the sample through a standardized SOD pressure viscometer at controlled temperature and pressure.

Features and Benefits
• U.S. Steel Method; ASTM Draft Method

• Test temperatures as low as –30°F (–34.4°C)


Click on the image to view a larger photo of the instrument.

K22680 Grease Mobility Test Apparatus
Performs grease mobility tests at low temperatures to predict pumpability characteristics. Determines the suitability of greases for applications in centralized or bulk systems where pumps, valves or pipes are used to distribute or transfer grease. Consists of pressure viscometer, cooling bath and refrigeration system. The stainless steel pressure viscometer is fitted with a modified No.1, 40:1 ratio capillary. After the sample is loaded in the pressure viscometer, the assembly is installed in the cooling bath and allowed to reach the test temperature. Mechanically refrigerated cooling bath can attain test temperatures as low as –30°F (–34.4°C) with stability of ±2°F (±1°C). With the sample at the test temperature, the flow of grease is started under the selected pressure on a nitrogen tank regulator. Flow per second is determined by collecting the grease for a specified period. Includes sample collector turntable.

Specifications
Conforms to: U.S. Steel Method; ASTM Draft Method

Minimum Temperature: –30°F (–34.4°C)

Control Stability: ±2°F (±1°C)

Electrical Requirements
115V 60Hz, Single Phase, 6A
220-240V 50 or 60Hz, Single Phase, 3A

Included Accessories
Grease Cylinder (pressure viscometer) with modified No.1, 40:1 capillary
Sample Collector Turntable

Dimensions lxwxh,in.(cm)
Cooling Chamber: 12x12x30 (30.5x30.5x76)
Refrigeration Unit: 15x12x12 (38x30x30)
Net Weight: 114 lbs (51.7kg)

Shipping Information
Shipping Weight: 188 lbs (85.3kg)
Dimensions: 18.4 Cu. ft.

Ordering Information
Catalog
Number
Product Description
Electrical Requirements
K22680
Grease Mobility Appparatus
115V 60Hz
K22685
Grease Mobility Appparatus
220-240V 50Hz
K22686
Grease Mobility Appparatus
220-240V 60Hz

Accessories
Catalog
Number
Product Description
K22680-0-22
Grease Cylinder with plunger and fittings
K22680-0-16
Capillary
250-100-001
Thermometer dial type
Range: -100 to +100
°F, with 2°F subdivisions