K56100 Cigre Bath, Oxidation Stability Apparatus

Test Method
Oxidation stability is determined by exposing the sample to a measured oxygen flow at elevated temperature in the presence of metal catalysts.

Features and Benefits

• Conforms to IP specifications

• Twelve-sample testing capability

• Microprocessor programmable high accuracy temperature control


Click on the image to view a larger photo of the instrument.

K56100 Cigre Bath, Oxidation Stability Apparatus
Constant temperature aluminum block type bath for oxidation stability tests in accordance with the Institute of Petroleum (IP) testing methods. Accommodates twelve sets of oxidation and absorption tubes. Insulated block bath operates efficiently at temperatures of up to 200°C (392°F). Microprocessor PID control provides quick temperature stabilization without overshoot, and the bath is protected by a an over-temperature control circuit that interrupts power should block temperature exceed a programmed cut-off point. Dual LED displays provide actual and setpoint temperature values in °C/°F format. Communications software (RS232, etc.), ramp-to-set and other enhanced features are available as extra cost options. Contact your Koehler representative for information. A bank of twelve flowmeters on a movable stand regulates oxygen flow at 1 ± 0.1L/h to each oil sample per IP specifications. Includes soap bubble flowmeter for checking oxygen flow rate.

Specifications
Conforms to: IP 48, IP 280, IP 306, IP 307, IP 335

Testing Capacity: Twelve samples

Temperature Range: 80 to 200°C

Temperature Uniformity: ±0.2°C

Air Flow Control:
Standard Model: 1L/h to each sample
IP 48 Model: 15L/h to each sample

Electrical Requirements
115V 50/60Hz, Single Phase, 12.1A
220-240V 50/60Hz, Single Phase, 6.3A

Included Accessories
Soap Bubble Flowmeter

Dimensions
Bath: dia.xh,in.(cm)
17x22 (43.2x55.9)
Flowmeter Stand: lxwxh,in.(cm)
24x8x30.25 (61x20.3x76.8)
Net Weight: 186 lbs (84.4kg)

Shipping Information
Shipping Weight: 245 lbs (111.1kg)
Dimensions: 16.7 Cu. ft.

Ordering Information
Catalog
Number
Product Description
Electrical Requirements
K56100
Cigre Bath, Solid Block Bath
115V 50/60Hz
K56105 Cigre Bath, Liquid Bath
K56190
Cigre Bath, Solid Block Bath
220-240V 50/60Hz
K56195 Cigre Bath, Liquid Bath
K56200
Cigre Bath, Solid Block Bath for IP 48 Method
115V 50/60Hz
K56205 Cigre Bath, Liquid Bath for IP 48 Method
K56290 Cigre Bath, Solid Block Bath for IP 48 Method 220-240V 50/60Hz
K56295
Cigre Bath, Liquid Bath for IP 48 Method

Accessories
Catalog
Number
Product Description
K56110
Set of Glassware
Includes one each oxidation and absorption tube. For IP 48, IP 280, IP 306, IP 307, IP 335.
250-000-09C
ASTM 9C Thermometer
Range: -5 to +110
°C
Equivalent to IP 15C Thermometer.
250-000-41C
ASTM 41C Thermometer
Range: 98 to 152
°C
Equivalent to IP 81C Thermometer.
When inquiring about NIST traceable certified thermometers, please refer to the catalog number for the corresponding thermometer and replace the middle three zeros in the catalog number with 004. Example: 251-000-01C ASTM 1C Thermometer would be 251-004-01C ASTM 1C Certified Thermometer.

www.koehlerinstrument.com