Automatic Dielectric Breakdown Tester
The majority of high-voltage transformers, cables, switchgears, transducers, capacitors, and rectifiers use insulating oils for insulating electrically live parts and to carry off thermal energy. The quality of the insulating oil must be checked at regular intervals to ensure a long equipment service life. The most important requirement of an insulating oil is a high dielectric strength. Determination of the dielectric breakdown voltage of insulating oils provides an early detection method for any reduction in the insulating properties.
- Conforms to ASTM D877, D1816, IEC 60156 and related test specifications
- Output voltage: K16185: 75kV, K16186: 90kV ,K16187: 105kV
- Built-in printer offers direct evaluation and reporting of results
- Internal battery, external 12V power supply
- Automatic Vernier function for electrode gap spacing
- Measurement of silicone-based oils
- Internal temperature measurement of oil sample
- USB flash drive compatibility
- Measuring Cell
- Gap Clearance Setting Device
- Magnetic Oil Stirrer
- Magnetic Stirrer Remover
- Electrode Polishing Cloth
- Mushroom Electrode
- USB Flash Drive
- Power Cable
- Paper Roll
- Operation Manual
- Storage Box
Conforms to the specifications of: ASTM D877, D1816 ; IEC 60156
Output Voltage: Up to 75kV rms symmetrical
Voltage measurement accuracy:
• K16185: 75kV
• K16186: 90kV
• K16187: 105kV
Voltage slew rate:
•K16185: 0 - 75kV ± 1kV• K16186: 0 - 90kV ± 1kV
• K16187: 0 - 105kV ± 1kV
Resolution (displayed): 0.1kV
Operating temperature: 0 – 50°C
Storage Temperature: -20°C to + 60°C
Temperature measurement resolution: ±1°C
Power frequency: 48 – 63Hz
High voltage shutoff time after breakdown, μs: < 10 (4 typical)
Measuring cell volume, cm3: 500
WxHxD,in.(cm):
• K16185: 20.08x12x13.15” (51x30x33.4cm)
• K16186: 20.08x12x13.78” (51x30x35cm)
• K16187: 20.08x12x13.78” (51x30x35cm)
Net Weight: 66 lbs (22kg)